All Categories
Polarographic; self stirring
One touch(TM) control(TM)
Low maintenance
Thermo Orion Auto Stir BOD Probe with Calibration Chamber and Stand 10 cm/sec. For use with 860A meters, laboratory. A new level of speed, convenience, and accuracy is featured in this Orion BOD Probe from Thermo Scientific. The built in stirrer provides vigorous sample agitation, preventing oxygen stratification, and can easily be disassembled for cleaning. The probe stand, which is free standing, can be used to store your probe when not in use, and also functions as an air calibration beaker. Polarographic, Self stirring. One touch control. Low maintenance. Cable length: 2m. Connector: 13 pin DIN. Response time : 90% of final value in 10 sec, 95% of final value in 15 sec, 99% of final value in 60 sec. Minimum sample flow: 10 cm/sec. Oxygen consumption: 0.008µg/h (mg/L) -11 at 20°C. Maximum allowed overpressure: 6 bar. Sample temperature: 0 to 50°C. Sample temperature: 0 to 50°C. Temperature sensor: Dual thermally separated. Drift: approx. 0.1% per day. Electrolyte lifetime: 180 days. Storage temperature : -5 to 50°C. Membrane cap: 080515. Maintenance kit: 080513. Orion #: 086020A.