--This instrument is mainly used to determine the kinematic viscosity of Newtonian or near-Newtonian fluid coating can also be used in accordance with the need to compare measurements.
--The correction coefficient K value is the time after the filling of the viscosity cup with the secondary standard oil under the set temperature conditions (such as 20 °C ± 0.1 °C), the flow time should be in the range of 30s ~ 100s, the K value should be 1 ± Within 0.03 range. The viscometer should be periodically calibrated to the extent of its frequency of use.
--The viscosity cup is placed on a measuring cup ring with a horizontal bubble, and the horizontal screw of the base is adjusted to center the bubble. The capacity of the receiving cup is about 150 ml, which is used for measuring the liquid.
--This instrument is mainly used to determine the kinematic viscosity of Newtonian or near-Newtonian fluid coating can also be used in accordance with the need to compare measurements.
Principle Viscosity LND-1 viscometer is the viscosity of the conditions, i.e. a certain amount of the sample at a temperature and a predetermined aperture holes from flowing out of time, in seconds. The following formula can be converted with the sample flow time in seconds (s) into the kinematic viscosity values (mm2 / s). Y = (t-6.0) /0.223 (30s - 100s) Where, t - flow time (s), Y - kinematic viscosity (mm2 / s) Stock up time: 2 days you will receive it within 5-9 working days.