Pipe size Suitable pipe size: 15ml: Inner diameter: 16mm, outer diameter: 18.5mm, length: 87.5mm. Pipes over 87.5mm in length and 18.5mm in diameter shall not be used. 20ml: Inner diameter: 18mm, outer diameter: 20mm, length: 95mm. Pipes over 95mm in length and 20 mm in diameter shall not be used. 50ml: Inner diameter 30mm, outer diameter 32mm, length 96mm. Pipes over 96mm in length and 32mm in diameter shall not be used. 100ml: Inner diameter: 38mm, outer diameter: 40mm, length: 95mm. Pipes over 95mm in length and 40mm in diameter shall not be used. Suitable centrifugal pipe type ACD gel tubes, glass tubes and plastic pipes and other pipes. Parameters: Speed range: 0-4000r / min Timing range: 0-99min Relative centrifugal force range: 0-2100xg Rotor specification: 15ml x 12, 50ml x 6,100ml x 4, 10ml x 12 Machine power: 200W Display mode: LCD NOTES The amount of sample added to each centrifugal tube must be basically equal. The sample amount of the rotor centrifugal tube is not more than 80% of the centrifugal tube capacity. The reagent can be symmetrically inserted into the rotor with only a small number of test tubes, which achieves the effect of central symmetry, but asymmetric operation is absolutely not allowed. Operations Put centrifuge tubes in the motor,close the cover plate and turn on the power supply. Press the speed knob to start and tentatively work, rotate the speed knob to adjust the speed, rotate the timing knob to set the timing, and start the timing after the speed reaches the set value. The speed range is 0-4000 rpm, which can be adjusted accordingto actual needs. When using, the cover should be properly capped to reduce the resistance of noise air.