All Categories
Specification: Voltage: 110V 60Hz Power: 185W Motor: 22000rpm Homogenizer 1: 8000 ~ 18000r/min Homogenizer 2: 8000 ~ 20000r/min Single throughput: 10~1000ml Glassware capacity: 5ml,10ml,15ml,20ml Continuously working: <5mins Ambient temperature: -10 ~ 45℃ Attention: The instrument should be placed on a horizontal work surface. The homogenizing tool required for installation must be in place. When the power supply test is turned on, use a cup or vessel containing water to flood the plum blossom head in the water and turn on the power switch. To quickly adjust the speed knob, there is no obvious metal friction sound then work Cut the plants or microorganisms to be tested into small pieces or small pieces, place them in a glassware that is randomly placed, and sandwich them in the middle of the magnet holder (seven-shaped shape), loosen the 8M/M fixing screws of the fixed motor base, and slowly align them. Move the vessel mouth down so that the homogenizing head is 2-3mm from the bottom of the vessel, and then tighten the 8M/M screw Instrument maintenance: After long-term use, ptfe in the high-speed dispersing tool is a fragile part, please replace it Note: after each work, remove the cutter and rinse it with water to dry